2B Technologies 臭氧分析仪Models 106-L, 106-M, 106-MH
Ozone Monitor Models 106-L, 106-M, 106-MH and 106-H
For Industrial Ozone Applications
For industrial ozone applications, 2B Tech offers the Model 106-L, 106-M, 106-MH and 106-H Ozone Monitors, where L, M, MH and H refer to Low (0-100 ppm), Medium (0-1,000 ppm)06-, Medium-High (0-10,000 ppm) and High (0-20 wt%) ozone concentrations, respectively. These instruments have different optical path lengths in order to accommodate an extremely wide range of ozone concentrations, spanning more than eight orders of magnitude. Additionally, the Model 106-H was designed with a flow-through path that can be pressurized to up to 50 psi for in-line measurements with ozone generators. All four models are available in either a standard enclosure or Industrial enclosure. Configurations for multi-channel sampling are available for Models 106-L, 106-M, and 106-MH.
The Model 106 series was designed as an "ozone monitor on a board" in which nearly all of the components are mounted directly to the printed circuit board. Thus, the Model 106 may be purchased without the enclosure for OEM applications. See: Model OEM-106.
The Model 106-L is designed for ambient low measurements down to very low ppb levels and has been designated by the EPA as a Federal E Method (FEM) for monitoring for compliance with the US Clean Air Act: EQOA–0914–218.
Our video describes the different models and configurations for our Model 106-Series Ozone Monitors, and could help determine which model is best for a particular application. Click here or scroll down on this page to view the video.
Specifications for Different Optical Benches
106-L | 0-100 ppm | 0.0001 ppm (0.1 ppb) | Higher of 0.0015 ppm (1.5 ppb) or 2% of reading |
106-M | 0-1,000 ppm | 0.01 ppm | Higher of 0.01 ppm or 2% of reading |
106-MH | 0-10,000 ppm (0-1 vol%) | 0.1 ppm | Higher of 0.05 ppm or 2% of reading |
106-H | 0-20 wt% O2 or Air | 0.01 wt% | Higher of 0.01 wt% or 2% of reading |
General Specifications
Measurement Principle | UV Absorption at 254 nm |
Measurement Interval | 2 s |
Data Averaging Options | 2 s, 10 s, 1 min, 5 min, 1 hr |
Flow Rate | ~1 Liter/min; Up to 20 L/min for Model 106-H |
Data Storage | 32,736 lines (10 s avg. = 3.7 days; 5 min avg = 113 days) |
Choice of Units | ppb, pphm, ppm, µg m-3, mg m-3 (for Models 106-L and 106-M) ppm, mg m-3 (for Model 106-MH) vol%, wt% air, wt% O2, g/m3, g/Nm3 (for Model 106-H) |
Data Outputs | USB, RS232, 0-2.5 V Analog, 4-20 mA, LCD Display |
Power Requirements | 100-240 V AC, 50/60 Hz 11-28 V DC, nominally 500 mA at 12 V DC, 6 watt for Models 106-L, M, and MH 11-28 V DC, nominally 325 mA at 12 V DC, 3.9 watt for Model 106-H |
Size (standard enclosure) | 3.6 x 7.9 x 9.4 inches (9 x 20 x 24 cm) |
Weight (standard and OEM) | 4.2 lb (1.9 kg) for Model 106-L; 2.8 lb (1.3 kg) for OEM-106-L 3.9 lb (1.8 kg) for Models 106-M, -MH; 2.5 lb (1.1 kg) for OEM-106-M, -MH 4.4 lb (2.0 kg) for Model 106-H; 2.9 lb (1.3 kg) for OEM-106-H |
Size (Industrial enclosure) | 16 x 14 x 9.3 inches (40.7 x 35.7 x 23.6 cm) H x W x D |
Weight (Industrial enclosure) | ~14 lb (6.4 kg) depending on model and multichannel options |
Features
Measurement based on UV absorption
Low power consumption; can be battery operated (optional lithium-ion battery)
Internal data logger with real-time clock
2-s measurement interval
On-board microprocessor with interactive menus includes data averaging options of 10 s, 1 m, 5 m and 1 hr
USB and RS-232 output of time/date, O3 concentration, internal temperature and pressure
Analog output (0-2.5 V and 4-20 mA) of ozone concentration in user selected units
Two 2-level relays for control purposes (e.g., control of ozone source or turn warning light on and off); Four 2-level relays in Industrial enclosure models and in the 3-channel and 6-channel optional configurations
Long-life pump (15,000 hr) for Models 106-L, 106-M and 106-MH
Bluetooth option for wireless data transmission
Optional configurations for 3-channel or 6-channel air sampling for Models 106-L, 106-M, and 106-MH
- 供应 buehler-technologies 温度传感器 TT-77
- 供应Heat and Sensor Technologies 辐射加热板 P2436AX0
- 供应 Viking Technologies 电容器 154-30
- 施耐德电气携手Hailo Technologies,以AI技术赋能工业自动化
- 日本机械展3D printing technologies EXPO
- 罗德与施瓦茨携手Skylo Technologies增强非地面网络(NTN)测试服务
- AvePoint Japan|日本教育展EDIX出展社|H3C Japan Technologies
- Class Technologies Inc.|日本教育展EDIX出展社|Google for Education
- NK technologies湿度传感器AS0-NCAC-120
- 宝飾輸入卸業販売|国際宝飾展出展社|Buysell Technologies
联系方式
- 地址:长沙 长沙市芙蓉区远大一路408号中栋406房
- 邮编:410005
- 电话:0731-89820291
- 销售经理:刘意
- 手机:18974821365
- 传真:86-0731-89820291
- QQ:953952240
- Email:lypeper@163.com